MRI SAFETY INFORMATION - RemeOs™
Non-clinical testing demonstrated that the RemeOs™ Screw is MR Conditional. A patient with this device can be scanned safely in an MR system under the following conditions. Failure to follow these conditions may result in injury:
Static Magnetic Field Strength (B0) |
1.5T or 3.0T |
RF Excitation | Circularly Polarized (CP) |
RF Transmit Coil Type | Volume RF body coil |
Operating Mode | Normal Operating Mode |
Maximum Whole-Body SAR |
2 W/kg (Normal Operating Mode) |
Maximum Head SAR | 3.2W/kg |
Scan Duration | Under the scan conditions defined, the implant can be scanned continuously for 25 minutes. With 5 minutes break, this can be repeated twice in 60 minutes. |